HK-8080 Residual Chlorine Analyzer

It can be used to monitor the residual chlorine online in the environment of tap water, pipe network water, water treatment disinfection and wastewater treatment.




Supplies category:Online Environment Protection Analyzer

Material code:HK-8080

Manufacturers,:HKY TECHNOLOGY CO.,LTD.



  • Introduction
  • Features
  • Specifications

HK-8080 Residual Chlorine Analyzer adopts an electrochemical technology to realize real-time online monitoring of residual chlorine by measuring the relationship between the current change caused by chemical reaction on the electrode and the concentration of analyte. It can meet the online monitoring of four parameters at the same time: residual chlorine, HClO, temperature and pH. It can be used to monitor the residual chlorine online in the environment of tap water, pipe network water, water treatment disinfection and wastewater treatment.

 Optional pH automatic or manual compensation

 Unique temperature compensation method can eliminate the influence of temperature on measurement results.

 Realize monitoring at the same time of four parameters : residual chlorine, HClO, temperature and pH.

 The electrochemical method without reagent.

 Equipped with flow cell of manual sampling, and the manual sampling is convenient and quick.

 Real-time online monitoring without reagent replacement.

 SD card can be used to upgrade on site or retrieve data.

Measurement Range

Residual chlorine: 0.0020.00mg/L

HOCl: 0.0010.00mg/L

pH: 0.0014.00

Temperature: 0.050.0 ℃

Resolution

Residual chlorine: 0.01mg/L

HOCl: 0.01 mg/L

pH: 0.01

Temperature: 0.1℃

Accuracy

Residual chlorine: Better than ± 2% or ± 0.02mg/l, which one is bigger

HOCl: Better than ± 2% or ± 0.02mg/l, which one is bigger

pH:±0.05

Temperature:±0.5℃

Min. Detection Limit

0.01mg/L

Response Time (T90)

≤90s

Calibration Method

Laboratory comparison

PH Compensation Mode

Manual / Auto

Temp Compensation Mode

Manual / Auto

Sample Requirement

Temperature: (550)

Flow: 3040 L/h

pH: 49

Solid contents in sample: ≤5μm (no gel and grease)

Output Current

(020)mA or (420)mA (Two Isolated Outputs)

Output Current Accuracy

±0.1mA

Max. Load Resistance

800Ω (20mA max.)

Environment Requirement

Environment Temperature: (545)

Environment humidity: ≤90% (Without condensation)

Power Supply

AC(100240)V, (5060)Hz

Storage Temperature

-2055 (above 0for electrode)

Communication

Internal Communication (without wire)

Optional Items

RS485 or Profibus (choose one), HART

Alarm Output

Dual normally open programmable relay contacts: DC 5A/30V or AC 5A/250V

Relevant recommended products

×

Ask for your quotation

HK-8080 Residual Chlorine Analyzer

Your message

Please try to complete your information so that we can deal with your problem more quickly and efficiently

  • The name*

  • gender*

  • email*

  • Telephone (please do not forget to fill in)*

  • Name of the entity*

  • address*

  • 城市

  • The zip code*

  • countries*

  • provinces*

Your requirements will be transferred to relevant personnel for processing. Clicking "send" indicates that you agree with mettler Toledo to contact you. We will try to meet your requirements and communicate with you about our products and services.
×