HK-5200 Series Thermostatic Control Device is an important necessary equipment of SWAS which should be mounted on the side of low temperature instrument’s rack. The device controls the sample stream at around 25oC that can ensure all of on-line instruments measurement accurate and power plant production stable.
The refrigeration copper tubing of the device adopts coil distribution in evaporator, so does the sample stream tubing. The two kinds of tubing are isolated each other, so that the temperature inside of evaporator is well-distributed, ensure thermostatic control accuracy. The sample stream tubing uses compression fitting for process connection which is convenient to do maintenance.
The refrigeration copper tubing of the device adopts coil distribution in evaporator, so does the sample stream tubing. The two kinds of tubing are isolated each other, so that the temperature inside of evaporator is well-distributed, ensure thermostatic control accuracy. The sample stream tubing uses compression fitting for process connection which is convenient to do maintenance.
. HIGH-POWER, WATER-COOLED COMPRESSOR—Refrigeration system has over-voltage and over-flow protection.
. UTILIZE HIGH-EFFICIENCY VORTEX COOLER TECHNOLOGY—Cool sample stream again, the flow of cooling water adjustable that ensure outlet temperature of each sample stream at around 25oC
. AUTO CONTROL FUNCTION—Mount control system and mechanical system separately, with temperature display, device operation completed by control system
. CUSTOMIZED SOLUTION—Design system as per power plant requirements
. UTILIZE HIGH-EFFICIENCY VORTEX COOLER TECHNOLOGY—Cool sample stream again, the flow of cooling water adjustable that ensure outlet temperature of each sample stream at around 25oC
. AUTO CONTROL FUNCTION—Mount control system and mechanical system separately, with temperature display, device operation completed by control system
. CUSTOMIZED SOLUTION—Design system as per power plant requirements
Model
|
HK-5210
|
HK-5220
|
HK-5230
|
||
Constant temperature Points N
|
≤3
|
4~11
|
≥12
|
||
Sample temperature at inlet
|
≤40℃
|
≤40℃
|
≤40℃
|
||
Sample temperature at outlet
|
(25±1)℃
|
(25±1)℃
|
(25±1)℃
|
||
The flux of the constant temperature sample
|
900mL
|
3600mL
|
7200mL
|
||
Cooling capacity KW
|
3
|
12
|
24
|
||
Refrigeration power Hp |
≤1.5 |
≤5 |
≤10 |
||
Power supply
|
3
|
5
|
10
|
||
Circulation pump’s power KW
|
0.8
|
1.5
|
22
|
||
Cooling water requirements
|
P≤0.3MPa
Q≤1 t/h
|
P≤0.3MPa
Q≤1.8 t/h
|
P≤0.3MPa
Q≤3.6 t/h
|